Hemocompatibility: Complement Activation, SC5b-9 (GLP) - Internationally Sourced

Complement activation asks whether your device triggers the immune cascade that a foreign surface in the bloodstream can set off. Human serum is exposed to the device and SC5b-9 generation is measured by ELISA against controls.

SKU: GRK-4006-04X
Availability:
Price: Call for pricing

When this test applies

Required wherever the device contacts circulating blood, under ISO 10993-1:2025. ISO 10993-4 does not prescribe one test — it sets out which blood-interaction categories apply to which device type, and complement activation is one of them. It matters most for devices with large blood-contacting surface areas and extended contact — oxygenators, dialysers, circuits — where cascade activation has clinical consequences.

When to run it

After cytotoxicity, and once the material set and surface finish are settled. Blood interaction is sensitive to surface topography and to processing residues, so running it on a material you still intend to change tends to buy a result you have to repeat.

What a pass tells you

That SC5b-9 generation is not significantly greater than the concurrent negative control. Worth knowing before you read a report: ISO 10993-4 sets no fixed numeric threshold here, so the conclusion rests on the comparison to controls rather than on a published limit.

What it does not tell you

Complement is one arm of the immune and coagulation response. It says nothing about hemolysis, platelet consumption or clotting time, and a device can activate complement without being hemolytic. ISO 10993-4 may call for several of the blood-interaction tests together, and a pass on one says nothing about the others.

Method

Testing per ISO 10993-4:2017. Human serum is exposed to the device or a representative sample and SC5b-9 is quantified by ELISA against concurrent negative and positive controls. Where a representative sample rather than the whole device is used, extraction ratios follow ISO 10993-12: 6 cm²/mL for material under 0.5 mm thick, 3 cm²/mL at 0.5 mm or thicker, or 0.1 to 0.2 g/mL where surface area is indeterminate.

Where it runs

This method is internationally sourced. The laboratory is ISO/IEC 17025 accredited and GLP certified, and the report is issued under Groenakker cover, in Groenakker’s report template or in your document template on request at no extra cost.

Usually ordered alongside

Platelet and leukocyte counts and PTT, which together with this test cover the thrombosis and immune categories ISO 10993-4 defines. Hemolysis is normally run as well.

Products specifications
SubcontractedYes
StandardsISO 10993-4:2017
EndpointHemocompatibility
Method detail
Test systemIn vitro - human serum
MethodComplement activation, SC5b-9 by ELISA
ExtractionDevice or extract incubated with human serum. Extraction ratio per ISO 10993-12: 6 cm2/mL for material under 0.5 mm thick, 3 cm2/mL at 0.5 mm or thicker, or 0.1-0.2 g/mL where surface area is indeterminate.
Acceptance criteriaPer ISO 10993-4, SC5b-9 generation is assessed against concurrent negative and positive controls; the device passes when activation is not significantly greater than the negative control. The standard sets no fixed numeric threshold.
DeliverableGLP-compliant final report with raw data, issued in Groenakker’s report template or in your document template on request at no extra cost, under Groenakker’s ISO 13485:2016 certified quality system.
Testing locationInternational (OUS)
*
*
*